ZEISS Originals CMM ACCURA II 9/16/8 Immagine del prodotto Front View L
ZEISS Originals CMM ACCURA II 9/16/8 Immagine del prodotto

CMM Usate ZEISS Originals

ZEISS Originals CMM ACCURA II 9/16/8

600099-2830-056
Su richiesta
Area di misura X
900
Area di misura Y
1600
Area di misura Z
800
Software
Software

Le CMM e gli strumenti di misura usati, se controllati e ricondizionati secondo i rigorosi criteri della linea ZEISS Originals, possono offrire prestazioni eccellenti pari a unità di nuova produzione, consentendo di ridurre sensibilmente il budget di spesa. Se siete interessati... non esitate a contattarci! NOTA: il prezzo indicato non comprende costi di trasporto, installazione e formazione.

Nota sul prodotto

CMM ACCURA II 9/16/8

construction year: 2018 refurbished: 2023

Coordinate measuring machines with the ZEISS Originals seal are refurbished used devices with components that may be made of granite.

Granite is a natural product and shows signs of wear (scratches, etc.). These signs of use do not constitute a defect that justifies a complaint. The promised measuring accuracy of the coordinate measuring machine is not affected by the traces of use.

CMM ACCURA 9/16/8 ZEISS CNC Coordinate Measuring Machine (CMM) in bridge design

  • measuring volume X = 900 mm, Y = 1600 mm, Z = 800 mm

  • Table Height TH = 752 mm

  • incl. DotScan preparation

  • glass-ceramic scales in all 3 axes

  • integrated passive vibration damping Specifications here below hold for

  • S1: active sensors VAST XT gold, VAST gold, VAST XTR gold

  • S2: passive sensor VAST XXT with RDS

  • temperature range 20 - 26 °C, remaining constant for 48 hours

  • maximum temperature gradients 1.0 K/h 2.0 K/d 1.0 K/m Max. Permiss. length measurement Error (MPE) as per ISO 10360-2:2009

  • MPE_E0 (3D) = (1.2 + L/350) µm for S1

  • MPE_E0 (3D) = (1.6 + L/350) µm for S2 Max. Permiss. Error (MPE) of scanning probing as per ISO 10360-4:2000 within Maximum Permissible Time (MPT) for scanning test

  • MPE_THP = 2.0 µm in MPT_tau = 40 s for S1

  • MPE_THP = 2.5 µm in MPT_tau = 50 s for S2 Single stylus form error as per ISO 10360-5:2010

  • MPE_PFTU = 1.2 µm for S1

  • MPE_PFTU = 1.7 µm for S2 Form measuring tolerance as per ISO 12181:2011 (VDI/VDE 2617 part 2.2)

  • MPE_RONt (MZCI) = 1.4 µm for S1

  • MPE_RONt (MZCI) = 1.7 µm for S2

  • pre-wired for tactile active and tactile passive sensors, for Multi Application Sensor System (mass)

  • VAST navigator and/or VAST XXT measurement technologies (tactile active scanning and/or tactile passive scanning)